63 03 452 D3352
116
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.8 Checking the data paths
Installation Manual ORTHOPHOS XG 3D / Ceph
11.8
Checking the data paths
The 2D and 3D data paths can be checked by creating panoramic, ceph
and sensor (3D) test images. To create test images, proceed as follows:
✔
SIDEXIS XG must be open.
➢ In SIDEXIS XG, call the
"Test exposures/test images"
menu:
"Utilities"
/
"Constancy test..."
/
"3D"
/
"Select X-ray device"
/
"Service exposure "
/
"Select X-ray component"
/
"Test exposures/
test images"
11.8.1
Creating 2D test images
✔
The menu
"Test exposures/test images"
is called.
1. In the structure tree, under
"Test exposures 2D"
, select the
"2D test
image"
element.
The menu for creating the panoramic and ceph test image
appears in the action window.
2. Click on the
"Image acquisition Panorama"
or
"Image acquisition
Ceph"
button to create the appropriate test image.
The exposure window opens in SIDEXIS XG. It indicates the
current status of exposure readiness.
3. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal indicating the end of the exposure
(double beep) sounds (if it has been configured).
The test image appears in the exposure window.