11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.6 Performing a mechanical function test
båÖäáëÜ
Performing a 3D (volume) test cycle
✔
The unit must be switched on.
1. Touch the 3D symbol at the top of the touchscreen.
The exposure program for volume exposure (3D) is selected.
2. Press the T key.
On the touchscreen, the display of the kV/mA value, the
exposure time and the patient symbols is hidden. Two test cycle
symbols appear.
3. Press the release button.
The test cycle is started.
4. Wait until the test cycle has been completed.
Quitting the test mode
➢ Press the T key again.
The program exits test cycle mode.
T
T