13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
3. Overwrite the adjustment values for S1, S2, and S3 in the text boxes
of the
"Ceph - Fixed point of rotation"
menu with the measured
values. For information about the shifting direction (input of +/- sign in
the menu), see the section entitled "Shifting direction of the exposed
image area". Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Ceph - Fixed point of rotation