10 Installation: Ceph arm
Sirona Dental Systems GmbH
10.3 Installation
Installation Manual ORTHOPHOS XG 3D / Ceph
9. Use the arrow keys (I) on the control panel to lower the unit until the
bearing bolts on the unit are at the same height as the drillings on the
ceph support arm.
NOTICE! Check the room height before you raise the unit. If the room
height is less than 2.27m (89 3/8") or 2.30m (90 1/2") for installation
with floor stand, you must limit the maximum travel height (S018: Test
step 2).
Installing the ceph arm
1. Fit the support arm onto the four bearing bolts.
2. Screw in the conical nuts (J) and tighten them firmly.
3. Detach the cables from the holder. Lay the cables over the support
arm in the groove between the slide and the support arm (from left to
right toward the opening (K)).
4x
J
IMPORTANT
The support arm may get tilted while you are tightening the conical nuts.
You can avoid this by lifting and moving the support arm gently while you
are tightening the nuts.