13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
7. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT:If automatic adjustment fails to produce values equal to
zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 231].
8. If adjustment is OK (the adjustment values in the text boxes are equal
to zero), save the values. [ → 206]
9. Continue with the next stage of the adjustment procedure [ → 234].
13.2.1.5.2
Manual adjustment: Ceph - Fixed point of rotation
The manual adjustment procedure is similar to the one for automatic
adjustment. The only difference is that the default adjustment values
calculated automatically by SIDEXIS XG in the
"Ceph - Fixed point of
rotation"
menu are overwritten.
1. Start the
"Ceph - Fixed point of rotation"
adjustment procedure as
described in the chapter entitled "Automatic adjustment: Ceph - Fixed
point of rotation (standard) [ → 228]".
A
Adjustment OK
● The exposed diaphragm area must lie
centered and straight in the image
section as well as inside the
superimposed auxiliary lines.
● A surrounding white border must be
present.
● The maximum density must lie in the
center of the diaphragm area.
B
Adjustment not OK