13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.1 General information about unit adjustment and calibration
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1.2.2.1
Navigation area
The navigation area contains a structure tree similar to the one you will
be familiar with from your Windows interface. The structure tree contains
all stages of the adjustment and calibration procedure you need to
complete in order to adjust and calibrate your system. The order in which
the elements appear in the structure tree determines the chronological
order of the procedure to adjust and calibrate the unit:
2D heading
2D adjustment
2D adjustment XG 5 / 3D / 3D ready
● Pan sensor adjustment
● Pan - Diaphragm
● Pan - Symmetry
● Ceph - Primary diaphragm
● Ceph - Fixed point of rotation
● Ceph - Main X-ray beam direction
● Ceph - Fixed point of rotation for Quickshot
● Ear plug alignment
3D adjustment
3D adjustment/calibration
● Sensor
● Diaphragm
● Shading
● Shading (5x5)*
(unit SW version V04.07.00 and higher)
● Geometry
(up to unit SW version V04.04.00)
– Center of rotation 1
– Center of rotation 2
– Center of rotation 3
– Center of rotation 4
– Center of rotation 5