12 Startup, measurements and tests (for USA/Canada only)
Sirona Dental Systems GmbH
12.7 Checking the tube current
Installation Manual ORTHOPHOS XG 3D / Ceph
2. Loosen the screw (D) and remove the lid of the tube assembly cover
(E).
3. DANGER! After switching off the unit, wait at least 4 minutes before
removing the measuring wires or reinserting the jumper!
Remove the measuring wires and bridge with the test points MA+/
MA- on the DX6 board again with the jumper (H).
4. Reattach the cover plate (G) to the tube assembly with the 4 screws
(F).
5. Reattach the lid of the tube assembly cover (E) to the unit and secure
it with the 3 screws (D).
E
D
F
G
E
D
D
D