13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
Resetting 3D adjustment
1. To reset 3D adjustment, click on the
"Reset adjustment"
element in
the structure tree under
"3D Reset adjustment"
.
2. To reset the adjustment values, click the corresponding button.
After resetting the values
Once the adjustment/calibration values have been reset, the unit must be
readjusted/recalibrated.
IMPORTANT
With the
"3D Reset adjustment"
function, only the
"Sensor "
and
"Diaphragm"
adjustments are reset. The calibration values from the
"Shading"
and
"Geometry"
menus are retained.