63 03 452 D3352
D3352
.
031.05.08
.
02 10.2015
257
Sirona Dental Systems GmbH
13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
13.2.4
Resetting adjustment/calibration
In an absolute emergency, the settings made for adjustment and
calibration can be reset in full or in part to the factory settings and/or
modified manually in the
"Adjustment/calibration"
menu.
Reset panoramic adjustment
1. To reset panoramic adjustment, click on the
"Panorama Reset
adjustment"
element in the structure tree under
"2D Reset
adjustment"
.
2. To reset the adjustment values, click on the corresponding button.
Reset ceph adjustment
1. To reset ceph adjustment, click on the
"Ceph - Reset adjustment"
element in the structure tree under
"2D Reset adjustment"
.
2. To reset the adjustment values, click the corresponding button.
CAUTION
You must make a note of the values displayed in the text boxes of the
menus before changing them. If necessary, you can then reset the
adjustment values to the factory settings.
Contact the Sirona Customer Service Center for more information (or to
enable the menu).