63 03 452 D3352
152
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.4
S018: Service for height adjustment
S018 XG 3D / XG 3D ready
* SR=service routine, ** SHZ=security access
SR*
SHZ**
Function
S018
Service for height adjustment
S018.2
No
Setting the maximum travel height
S018.3
No
Undoing the maximum travel height setting
S018.4
No
Checking the height adjustment sensor system
S018.5
No
Setting the minimum travel height
S018.6
No
Undoing the minimum travel height setting
S018.7
No
Setting the nominal angle for the occlusal bite block (as required by the
customer) (occlusal bite block only)
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!
S018.8
No
Function test of the occlusal bite block (occlusal bite block only)
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!
S018.9
No
Adjustment of the AD acceptance angle values and the zero point
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!
S018.10
No
Setting the PWM values for the height adjustment creep speed
(PWM=Pulse Width Modulation)
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!