13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.4 Mechanical adjustments
båÖäáëÜ
13.4
Mechanical adjustments
13.4.1
Mechanical adjustment: Ceph secondary diaphragm
1. Loosen screw (A) and remove the cover of the secondary diaphragm
by pulling it downward.
2. Loosen screws (B) slightly (approx. 2-3 turns).