63 03 452 D3352
D3352
.
031.05.08
.
02 10.2015
155
Sirona Dental Systems GmbH
11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.4.3
S018: Test step 4
Check the height adjustment sensor system
This service routine is used to move the unit up or down as far as the limit
switches using the Up/Down keys on the control panel. The "soft limit
positions" set by the software are ignored in this case.
S018_4 XG 3D / 3D ready
1. Call service routine S018.4 [ → 125].
Once the service routine has been selected, the current height
position is displayed in selection field 1.
Patient symbol keys 1 to 3 (F) show the switching state of the limit
switches (see table)
If the patient symbol key is lit, the corresponding switch is
activated, i.e. the unit is at a position value greater than 1500.
2. Use the UP/DOWN keys on the control panel to move the unit up and
down and use the patient symbol keys (F) to check the switching
states.
3. Exit the service routine [ → 129].
Display on the control
panel
Status
Meaning
Patient symbol key 1
lit
Correction switch activated
not lit
Correction switch not
activated
Patient symbol key 2
lit
Lower limit switch activated
not lit
Lower limit switch not
activated
Patient symbol key 3
lit
Upper limit switch activated
not lit
Upper limit switch not
activated