13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.1 General information about unit adjustment and calibration
Installation Manual ORTHOPHOS XG 3D / Ceph
Ceph - Fixed point of rotation
Adjustment of the fixed point of rotation can usually be performed directly
via precision adjustment. Only in exceptional cases, e.g. if the exposed
image area is completely outside the image section in an image acquired
with the
"Precision adjustment"
setting (E), is it necessary to perform
coarse adjustment prior to precision adjustment (F).
On the exposure with coarse adjustment (F), the exposed area is still
visible in the image field. Even in this extreme case, an adjustment would
still be possible.