13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.1 General information about unit adjustment and calibration
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1.6.3
Geometry phantom for volume calibration
* For enhanced representation, the unit is shown in the left-hand image
without combisensor.
Calibration of the XG3D
For sensor 3D adjustment, as well as for geometry 3D calibration, you
must insert the geometry phantom (A) into the pan bite block holder (B)
on the unit.
For diaphragm 3D adjustment, as well as for shading 3D calibration, you
must take the geometry phantom back out of the bite block holder.
NOTICE
Risk of damage to unit
Make absolutely sure that the geometry phantom is inserted in the bite
block holder of the unit in such a way that the body of the phantom and
the screw (C) point
away from the unit in the direction of the patient
(see
figure). Otherwise, this can result in a collision between the combisensor
and the geometry phantom and thereby damage the unit.
*
IMPORTANT
Once it has been inserted into the bite block holder, the geometry
phantom must be aligned vertically and horizontally with the spirit level,
so that calibration can be performed correctly.
A
B
C