13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
2. Once you have taken the exposure, measure distances Sx and Sy
with the SIDEXIS XG measuring ruler.
Tip: To facilitate the measuring procedure, you can also color the
image in SIDEXIS XG (see the SIDEXIS Operator's Manual).
3. Overwrite the adjustment values for Sx and Sy in the text boxes of the
"Pan - Diaphragm"
menu with the measured values.
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area". Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter
entitled "Automatic adjustment: Pan diaphragm (standard) [ → 216]".