13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
Horizontal correction
1. Loosen the screws (A) slightly (approx. 2 - 3 turns).
Tip: Do not unscrew them fully!
2. Adjust the ear plugs in the horizontal direction by turning the screw
(B) counterclockwise or clockwise. The direction of rotation for left-
arm (C) and right-arm (D) versions is shown in the drawing below.
3. Retighten screws (A) firmly.
Vertical correction
1. Turn the rotary table counterclockwise approx. 100 degrees until you
can see the screw (E) through the opening in the cover plate.
2. Loosen the screws (E) slightly (approx. 2 - 3 turns).
Tip: Do not unscrew them fully!