63 03 452 D3352
194
D3352
.
031.05.08
.
02 10.2015
13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.1 General information about unit adjustment and calibration
Installation Manual ORTHOPHOS XG 3D / Ceph
Adjusting the height of the unit during calibration: XG
As of unit software version V 03.04.00, you can even adjust the height of
the unit during the calibration procedure. If the unit is ready for an
exposure (after the
"Image acquisition"
button has been pressed in the
SIDEXIS XG
"Adjustment/calibration"
menu), the corresponding service
routine (S002.6/S010.10-14/S011.8, and S030.5) is displayed on the
control panel. All of these service routines allow the height adjustment
menu to be opened by pressing the Test key. The current unit height is
displayed in selection field 1 in this menu. You can then set the unit to the
desired height using the UP/DOWN keys on the control panel.
Press the Service key or the double arrow key to exit the height
adjustment menu.
13.1
General information about unit adjustment and
calibration
Please adhere to the following order when adjusting and calibrating the
system:
XG 3D calibration procedure
2D adjustment (panoramic)
● Pan sensor adjustment
● Pan - Diaphragm
● Pan - Symmetry
If a cephalometer is installed:
● Ceph - Primary diaphragm
● Ceph - Fixed point of rotation
● Ceph - Main X-ray beam direction
● Ceph - Fixed point of rotation for Quickshot
● Ear plug alignment
3D adjustment/calibration (volume)
● Sensor
● Aperture
● Shading
● Shading (5x5)*
● Geometry
● Dosimetry
* Unit SW version V04.07.00 and higher
Tip: It may be helpful to use the SIDEXIS XG coloring function to evaluate
the image.