63 03 452 D3352
D3352
.
031.05.08
.
02 10.2015
223
Sirona Dental Systems GmbH
13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
2. After taking the exposure, measure the distances S1, S2 and S3 with
the SIDEXIS XG measuring ruler.
If possible, measure in the lower area of the needles since they may
be slightly bent after several uses.
Always measure exactly from the center of the auxiliary line to the
center of the pin.
Tip: To facilitate the measuring procedure, you can also color the
image in SIDEXIS XG (see the SIDEXIS Operator's Manual).
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
of the
"Pan - Symmetrie"
menu with the measured values.
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area". Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter