63 03 452 D3352
222
D3352
.
031.05.08
.
02 10.2015
13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT:If automatic adjustment fails to produce values equal to
zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 222].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to zero), save the values. [ → 206]
6. Remove the OP XG needle phantom from the pan bite block holder
on the unit.
7. Continue with the next stage of the adjustment procedure. [ → 224]
13.2.1.3.2
Manual adjustment: Pan symmetry
The manual adjustment procedure is similar to the one for automatic
adjustment. The only difference is that the default adjustment values
calculated automatically by SIDEXIS XG in the
"Pan - Symmetrie"
menu
are overwritten by adjustment values calculated manually.
1. Start the
"Pan - Symmetrie"
adjustment procedure as described in
the chapter entitled "Automatic adjustment: Pan symmetry [ → 220]".
A
Adjustment OK
● The shadow of the center needle,
the needle image, and the
auxiliary lines must be coincident
and located one behind the other.
A tolerance (offset of needle from
the central auxiliary line) of ± 0.75
mm is admissible.
● Distance A1 must be 88.6 ± 1 mm.
● Distances A2 must be identical,
each being 44.3 ± 0.5 mm.
B
● A surrounding white border must
be present.
C
Adjustment not OK