63 03 452 D3352
172
D3352
.
031.05.08
.
02 10.2015
11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
5. Use arrow keys (E) to select the digit to be changed in selection field
1 ("Digit No. 12" in the example).
The corresponding patient symbol key (D) lights up.
Selection field 2 displays the value of the currently selected digit.
6. To change the value for the digit, use arrow keys (F) in selection field
2.
The Memory key (G) lights up.
7. Save the setting [ → 129].
8. Exit the service routine [ → 129].
9. Perform a restart of the device.
Содержание Orthophos XG 3D/Ceph