13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.3 Checking and adjusting the touchscreen
Installation Manual ORTHOPHOS XG 3D / Ceph
6. Touch the center of the green square in the bottom left corner of the
screen.
The fifth adjustment screen appears on the display.
7. Confirm the adjustment settings by clicking the
"Yes"
button.
The touchscreen is now adjusted.
To repeat the adjustment procedure, touch the
"No"
button.
To abort the adjustment procedure, touch the
"Abort"
button.