11 Initial startup
Sirona Dental Systems GmbH
11.9 Checking the unit configuration
Installation Manual ORTHOPHOS XG 3D / Ceph
11.9
Checking the unit configuration
Room and travel height
1. Check the room height before you raise the unit using the UP/DOWN
keys.
2. If the room height is less than 2.27 m (89 3/8") or 2.30 m (90 1/2") for
installation with the floor stand, you must limit the maximum travel
height.
Checking the hardware version: XG 3 / 5 / 3D / 3D ready
Hardware version
All units ordered with or without a ceph arm are preconfigured for this
version at the factory. You should, however, check the unit configuration
anyway.
➢ Select ceph mode on the control panel.
If ceph mode cannot be selected, even though a ceph arm is
installed, check the unit configuration using service routine S017.2.
Remote control
All units ordered with or without remote control are preconfigured for this
version at the factory. You should still check the unit configuration using
service routine S017.6.
Checking the occlusal bite block configuration: XG 3D / 3D ready
Occlusal bite block
All units ordered with or without occlusal bite block are preconfigured for
this version at the factory. You should still check the unit configuration
using service routine S017.18.
Set up the occlusal bite block in accordance with customer requirements
using service routines S018.7 through S018.10.