63 03 452 D3352
218
D3352
.
031.05.08
.
02 10.2015
13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
4. Repeat the procedure starting at Step 2 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT:If automatic adjustment fails to produce values equal to
zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 218].
5. If adjustment is OK (the adjustment values in the text boxes are equal
to zero), save the values. [ → 206]
6. Continue with the next stage of the adjustment procedure. [ → 220]
13.2.1.2.2
Manual adjustment: Pan diaphragm
The manual adjustment procedure is similar to the one for automatic
adjustment. The only difference is that the default adjustment values
calculated automatically by SIDEXIS XG in the
"Pan - Diaphragm"
menu
are overwritten by adjustment values calculated manually.
1. Start the
"Pan - Diaphragm"
adjustment procedure as described in
the chapter entitled "Automatic adjustment: Pan diaphragm
A
Adjustment OK
● The exposed diaphragm area must lie
horizontally centered in the image
section as well as inside the auxiliary
lines.
● A surrounding white border must be
present.
● The maximum density must lie in the
center of the diaphragm area.
B
Adjustment not OK