13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
3. Overwrite the adjustment values for S1, S2, S3, and S4 in the text
boxes of the
"Ceph - Primary diaphragm"
menu with the measured
values.
4. Continue with the adjustment procedure as described in the chapter