E51: Mixer not back in its vertical
position! Check if mixer is in the
vertical position and clear error if
OK!
Mixer could not come back in its
vertical position after mixing.
•
Mixer maybe hitting the
bottom transport belts.
•
Mixer is blocked.
•
Mixer motor is faulty.
•
Mixer motor driver on rack
board is faulty.
E54: Endpool full ! Remove all
racks from endpool and then clear
error!
This error will be cleared
automatically after removing the
racks from the end pool.
•
End pool full.
•
Check Rack out sensor.
E55: Rack lost in endpool!
Remove all racks from endpool
and then clear error!
Rack is released from gripper but
did not pass the End pool sensor.
•
Rack maybe on its side.
•
Check End pool full sensor.
•
Check conveyer belt motor.
E56: Checksum error motor
settings table!
Checksum error on motor settings
stored in Eeprom. Settings are
loaded from file. Check Motor
settings and save settings.
Possible hardware failure in
Eeprom.
E57: Checksum error current table! Checksum error on current settings
stored in Eeprom. Settings are
loaded from file. Check Current
settings and save settings.
Possible hardware failure in
Eeprom.
E58: Checksum error time-table!
Checksum error on timing settings
stored in Eeprom. Settings are
loaded from file. Check Timeout
settings and save settings.
Possible hardware failure in
Eeprom.
E59: Checksum error position-
table!
Checksum error on position
settings stored in Eeprom. Settings
are loaded from file. Check
Position settings and save settings.
Possible hardware failure in
Eeprom.
E60: Mixermotor error! (Hardware) Could not home mixer motor.
•
Check rack timing setting
T1.
•
Mixer motor is faulty
•
Mixer motor is blocked.
•
Mixer motor driver on rack
board is faulty.
•
Check the mixer motor
drives the gripper.(Loose belt,
loose pulleys )
Appendix for Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
191
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...