optimum sample speed. The sample tubes in the rack are faster available for the user but the
carousel shows more unused pipettes.
5.
Check for duplicate IDs
▪
Check for duplicate IDs
ON: It is not possible to run the same sample ID as long as this ID is in the
pipette carousel data buffer
▪
Check for duplicate IDs
OFF: It is possible to run the same sample ID, even when it is still stored in
the carousel data buffer.
6.
Virtual keyboard
▪
Virtual keyboard
ON: If keyboard input is required, a virtual keyboard automatically pops up on the
screen.
▪
Virtual keyboard
OFF: No pop-up screen of virtual keyboard.
7.
Print after measurement
▪ This will print the measured result on a single line directly to a dot matrix printer. If this option is
switched ON, it is also possible to print a new header at the top of the results.
When other printers then dot matrix are used, every result is printed on one new page.
8.
Sample probe
protection
▪
Sample probe
protection
OFF: The sample probe motor will push the sample probe down to the set
depth.
▪
Sample probe
protection
ON: The sample probe motor will stop when a certain current limit is
exceeded and returns to the home position.
▪ Reset the
Sample probe
protection
switch from ON to OFF and ON again.
The following numerical inputs can be made:
1.
Temperature in Celsius
. To set the correct room temperature.
2.
ESR sedimentation time in minutes
. To set the correct sedimentation time: 30 or 60 minutes. This time is
reset to default when Service mode is switched OFF.
3.
Pipette wash time (sec)
in seconds. Default value 7. This setting is reset to default when Service mode is
switched OFF.
4.
Pipette dry time (sec)
in seconds. Default value 5. This setting is reset to default when Service mode is
switched OFF.
5.
Sample probe depth
in millimeters.
6.
Mixing interval (30..90 sec)
time between two mixing operations of the rack in seconds. Default value 90.
Language
The selection box for the language is marked with a symbol depicting a globe.
Select the language for the software and the "Instructions For Use" by clicking on the appropriate language
name.
Note
: To switch to the selected language the first time, close and restart the software.
Starrsed program
80
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...