1. Separator errors.
2. Foam in the separator.
3. Waste pump cannot sufficiently remove the waste from the separator.
13.3 - Level 4 maintenance
The purpose of level 4 maintenance is to carry out the daily / weekly maintenance and replace the pump
tubing, bacterial filters and the Fill nozzle O-ring. After replacing those items, the instrument needs a Fill and
Clean sequence to clean the pipettes. Over a period protein builds up in the Westergren pipettes and needs
to be deproteinized using a strong cleaning agent.
Detailed instructions of this procedure can be found in the Work Instruction
.
13.3.1 - Pump tube replacement
New rinse pump tube assembly
ESRI090902
.
New saline pump tube assembly
ESRI090903
New tube replacement:
1. Open left cover.
2. Pull pump tube slightly downwards and at the same time towards the front of the unit to release the
tube out of the pump plate holder.
3. Remove the old tube from the peristaltic pump rotor.
4. Disconnect the tubing at both ends of the tube connectors.
5. Connect new tubing to both ends of the connectors.
6. Place one end of the tube in the pump plate holder.
7. Pull the new tube over the peristaltic pump rotor.
8. Pull pump tube slightly downwards and at the same time towards the back of the Starrsed RS.
If the tube is not fitted correctly or is worn the following symptoms can occur.
•
Liquid flowing back into the container.
•
First glass tube on the pipette belt is not washed sufficiently.
Note:
The wider bore tube is for the rinse pump.
The narrower bore tube is for the saline pump.
General maintenance
MRN-071-2-EN version 7.02 Starrsed RS
133
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...