15.17 - Fill nozzle
Normally the fill sequence takes about 3 seconds. However, if the fill sequence exceeds 10 seconds, a fill
time-out error will be generated. The ESR analyser unit aborts the fill sequence and this error message will
be shown on the display and reported.
Check for mechanical obstructions and remove them. If the error returns, call for service.
15.17.1 - Fill nozzle does not engage with pipette
1. Motor time out generated, fill nozzle stops half way up.
2. Check for mechanical obstruction.
3. Motor failure. Fatal error, call distributor.
15.17.2 - Fill nozzle not at fill position
E8 is the error code and in the status line the text Fill nozzle not in the fill position.
The fill nozzle did not reach the fill position in a certain time limit.
Possible reason;
•
Fill nozzle motor is faulty.
•
Fill nozzle motor driver is faulty.
•
Fill nozzle is blocked.
•
Fatal error, call distributor.
15.17.3 - Fill nozzle not at home position
E13 is the error code and in the status line the text Fill nozzle did not reach the Home position within a
certain time limit.
The fill nozzle did not reach the home top sensor with a certain time limit.
•
Fill nozzle motor faulty.
•
Fill nozzle motor driver is faulty.
•
Fill nozzle is blocked.
•
Fatal error call distributor.
15.18 - Piercing error
There are two error codes related to a piercing error: E7 and E20.
E7
Outer needle motor position error ! Timeout ! (piercing)
The Outer needle did not go down within a certain time limit. This can be caused by:
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
163
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...