Work instruction Number
230
Page
1 of 1
Purpose:
Preparation for transportation or
disposal of a Starrsed Instrument
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
002, April 2020
Decontamination of Starrsed RS
Prepare disinfectant Virkon (QRR 010046):.
Add
2
grams of disinfectant to
200 ml
deionised water.
(1% solution)
This disinfectant is only used for the fill and clean sequence of the Pipettes and for cleaning all external parts that
are exposed to blood.
1. Perform a "Fill and Clean" sequence with the pre-prepared disinfectant
2. Remove all reagent containers and replace with containers with DI water
3. Start "Wash all pipettes"
4. Prime the diluter 5 times
5. Disconnect and remove reagent containers
6. Perform "Prime all" until all liquid is removed from all tubing
7. Remove tubing from reagent pumps
8. Release tubing from pinch valves (Diluter valve, Sample valve and Vacuum valve)
9. Check if the separator is clean
10. Clean outside of the sample needle
11. Clean outside of instrument. Wipe outer surface and stainless steel plate below the pipettes with disinfectant.
12. Disconnect waste connection
13. Disconnect power
Preparation for temporary storage:
1. Cover machine with dust protection
2. Store the reagents at ambient temperature under dry conditions
Preparation for transport:
1. Cover machine with dust protection
2. Make the Starrsed RS ready for transportation
3. Send the Starrsed RS to the new address
Final preparation for permanent disposal:
1. Remove needle assembly and dispose of as biohazardous waste
2. Remove filter from separator and dispose of as biohazardous waste
3. Discard instrument according local regulations
Work instruction Starrsed RS
222
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...