Tube motor position error
▪
Blockage of the Outer needle
▪
Faulty Outer needle motor
▪
Faulty Tube motor driver on the needle board
E20
Outer needle motor position error ! (piercing error).
The Outer needle could not go down all the way. This can be caused by:
▪
Piercing position sensor was triggered
▪
Piercing position sensor was not triggered, check the piercing position sensor
▪
Tube 1 not aligned properly, check the tube 1 position setting and save the setting
For other error numbers and details See the
Appendix - Error list Starrsed RS
bij wijziging ook txt 18185 bekijken
15.19 - New rack adapter
Go to
Service screen
and select
Position motor
-
Select rack
and select the required rack type.
Go to
Test position
motor and
Go to TUBE
position and set in Tube # number 1.
The Gripper motor will move the rack to the first position of the rack. Check if the needle is above the first
hole. If the needle is not above the first hole turn the wheel on the motor until the needle is above the hole.
Check that the needle is going though the hole by using the Outer needle function
DOWN
.
Get the needle out of the hole press the function
UP
.
By turning the wheel the actual position numbers are changed.
When the needle is going nicely through the hole press the
Save as Tube 1 position
.
Next step is press Save settings.
Now all the new settings are saved in the file and Eeprom.
15.20 - Rack unit errors
The rack unit can give errors on the
•
•
•
15.20.1 - Gripper module
E61 till E67, E69, E71 E72, E80 till E88 are the error codes and the text in the status line begins with
"
Grabber did not reach the
":
Troubleshooting
164
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...