QC sample is not accepted and not performed
Starrsed Control ID is not known in LIMS.
Check barcode
QC result is not visible in QC History
A specific QC result cannot be found in the list of results. Check Lab-ID link
Deviating results
Systematic QC errors with a shift in control values (QC results are out of range)
The measured control values change abruptly up- or
downwards.
Do not compare 30 minute method with 60 minute
method result. The calculation method can give some
deviation in the general QC results statistics.
•
Check/clean instrument and perform a new QC
sample
•
If these errors persist perform maintenance step
•
Compare only results from one batch.
•
If Lab ID is used check the linked Starrsed Control
ID. It is possible that a new batch is in use without
changing to the new assayed mean value
Systematic QC errors with a trend in control values (QC results out of the range or nearly out of the range)
The measured control values change gradually upwards
or downwards.
•
Irregular or insufficient maintenance can cause
unnecessary QC errors and ESR errors/warnings
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
171
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...