Sample ID:
The sample identification (barcode) of the sample tube.
•
Dilution:
The dilution rate of this sample as calculated during the aspiration process.
•
Time filled:
The date and time when the sample was aspirated.
•
Time to go:
The number of minutes to wait until the sample will be measured.
The indicators at the right side show the current status of the selected pipette:
•
Pipette filled OK:
A sample has been aspirated into the pipette without problems.
•
Pipette dirty:
The sample has been measured and the pipette is marked to be washed when it reaches the rinse
station. This indicator is also on when a sample could not be aspirated properly.
•
Pipette disabled
In case of a failing, missing or broken pipette individual pipettes can be disabled, only when sample
mode is deactivated. Click on the blue circle of the pipette, click on the check box "Pipette disabled" and
confirm disabling.
Disabled pipettes are indicated with a light blue circle (nr. 21 on example screen).
During normal sampling of the instrument disabled pipettes will be skipped, also during End of Day
Wash, Fill & Clean, wash all pipettes and pipette flow test the disabled pipettes will be skipped.
Starrsed program
44
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...