When there is a sensor alarm, an alarm indicator is shown in the tab
Maintenance
.
All functions to check the status of the sensors are grouped under button
NOTE
: Clean sensors first before executing this function.
•
Check Fill stop sensor: Values must be within the following limits: FS 90..
140
..165
•
Check temperature sensor: Value must be equal to the actual room temperature near the pipette belt.
The value can be set in tab
Settings
.
•
Check Diluter start sensor: This sensor is only used in EDTA mode. If the diluter does not start during
the aspiration, the status of this sensor must be checked.
The value should be: Diluter start sensor 400-
550
-700.
•
Check Diluent flow sensor: This sensor is only used in
. When activated, the indicator
Stand-by is activated and the indicator Flow is not. When the button Test is clicked, the indicator Flow
must be shown as "activated". After finishing the test, both indicators must be shown as "activated".
•
Check Separator sensor: The value must be in range of <200
600
>700.
•
Check Flow sensor: The vacuum unit switches on and the values must be in this range:
Flow:
0980
± 60 Abs:
0320
±10
Note: If for example the yellow orifice is blocked the flow will be: Offset:
0050
±5.
•
NOTE: Clean sensors first before executing this function.
Check Measure sensor (In Service mode): When the sensor is not engaged with the pipette, the value
must be within the following limits: MS
50
±10.
Press the button
Measure
. The pipette currently at the measure position will be measured.
The results are displayed in graphical form. Raw data is also stored on the D: drive (D:\MeasureTest.txt).
Measure head start
position correct
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
73
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...