E61: Position motor error! Not on
position! (OUT)
Gripper did not reach the OUT
(release) position. Starting point:
FILL position.
Spindle is greased or oiled
•
Position (gripper) motor is
faulty or blocked
•
Motor driver on rack board is
faulty
•
Clean spindle thoroughly. Do
not lubricate!!
E62: Position motor error! Not on
position! (MIX)
Gripper did not reach the MIX
position. Starting point: OUT
position.
See E61
E63: Position motor error! Home
sensor!
Position motor did not reach the
home position or could not detect
home sensor.
•
Check position (gripper)
motor Home sensor.
•
Position (gripper) motor is
faulty.
•
Position (gripper) motor is
blocked.
•
Motor driver on rack board is
faulty.
•
Check position motor belt.
E64: Position motor error! Not on
position! (IN)
Gripper did not reach the IN
(catch) position. Starting point:
MIX position.
See E61
E65: Position motor error! Not on
position! (MIX)
Gripper did not reach the MIX
position. Starting point: FILL
position.
See E61
E66: Gripper error! Rack not in
gripper!
Follow the instructions on the
screen.
•
Check height of the tubes in
the rack.
•
Check gripper sensor.
•
Check rack adapter.
E67: Position motor error! Not on
position! (FILL)
Gripper did not reach the FILL
position. Starting point: MIX or
FILL position.
See E61
E68: Mixing motor current
overshoot. Check if the mixer is
blocked!
The mixer motor current exceeded
the Maximum mixing current
setting during mixing.
•
Check mixer motor settings.
•
Mixer motor is blocked.
•
Mixer motor is faulty
•
Mixer motor driver on rack
board is faulty.
•
Check mixer motor belt.
Appendix for Starrsed RS
192
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...