Selections are made by an ON-OFF switch in the screen and by numerical inputs.
If virtual keyboard is switched ON a virtual keyboard pops up for input the data. When Virtual keyboard is
switched OFF the arrow keys must be used for input of data.
When ready with press
Save settings
before you continue.
The following selections can be switched ON or OFF:
1.
30 min. method
▪
30 min. method
ON: ESR's are measured after 30 minutes.
▪
30 min. method
OFF: ESR's are measured after 60 minutes.
2.
EDTA mode
▪
EDTA mode
ON: Samples are presented in EDTA sample tubes. The samples are diluted in the
Starrsed RS.
▪
EDTA mode
OFF: Samples are presented in pre-diluted CITRATE sample tubes. Dilution on the
Starrsed RS is switched OFF.
3.
Temp. Correction
▪
Temp. Correction
ON: A temperature corrected value for the ESR is shown besides the actual
measured value.
▪
Temp. Correction
OFF: Only the actual measured ESR value is shown.
4.
Fast filling
▪
Fast filling
OFF (Default setting): The carousel is filled with optimum pipette usage but completion of
a sample rack takes a little longer. The rotation sequence of the carousel is evenly divided. In 60
minute method, the carousel moves one position every 40 seconds (approx.). In 30 minute method,
the carousel moves one position every 20 seconds (approx.).
Fast filling
ON: The carousel is filled with
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
79
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...