1. Loosen the screw which prevent the
outer needle to drop out of the
assembly.
2. Loosen the sample probe.
3. Pull the outer needle complete (with
sample probe) out the needle
assembly.
4. Mark each tube for easier
reconnecting to the correct nipple.
5. Disconnect the tubes from the outer
needle.
Needle exchange:
Install (new) sample probe ESRI050909 together with a new outer needle VERA059009
1. Slide the new sample probe into the (new) outer needle.
2. Make sure the Sample probe has a (new) O-ring QWLV050003.
3. Install (new) sample probe ESRI050909 together with the (new) outer needle
4. Put the sample probe in the outer needle.
Replace the needles onto the needle assembly:
1. Tighten the sample probe.
2. Replace the correct tubes on the outer needle.
3. Fasten the outer needle bolt.(Do not over-tighten the screw)
4. Replace the cover on the Rack unit.
13.10 - Cleaning Touch screen display
Any standard glass cleaner can be used to clean the touchscreen, but avoid products containing ammonia or
high concentrates (>70%) of alcohol or solvents.
Always spray the glass cleaner on a microfiber cloth or towel and then clean the touchscreen. Glass cleaner
sprayed directly on the monitor could possibly leak inside a non-sealed unit and cause damage.
Dirt and fingerprints do not affect the operation of a properly sealed touch display.
General maintenance
140
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...