The OUT (release) position. Starting point: FILL position
•
The MIX position. Starting point: OUT position
•
The HOME position. Could not reach or detect the home sensor
•
The IN (catch) position. Starting point: MIX position
•
The MIX position. Starting point FILL position
•
The MIX position. Starting point IN position
Causes and solutions
•
Transport gripper motor (VERA129014) is faulty
•
Motor driver on rack board is faulty
•
Check position (gripper) motor home sensor (VERA089012)
•
Position gripper motor (VERA129013) is faulty or blocked
•
Spindle is greased or oiled, clean the spindle
•
Check position motor belt
•
Check gripper sensor
•
Check height of the tubes in the rack
•
Check rack adapter
•
Check height position of the rack adapter
E73 till E75 are the error codes and in the status line the text begins with "
Position motor error!
":
•
is on the IN position. Could not send the rack to the IN pool to the first stop position.
•
is not on the OUT position. Could not release the rack from the gripper.
•
is not on the IN position. Could not send rack to the gripper.
Causes and solutions
•
Check the position motor
For more error numbers and details see the
15.20.2 - Mixer module
E60 and E90 is the error code and the text in the status line Mixer motor error. Could not home mixer motor
•
Check rack timing setting T1
•
Mixer motor is faulty.
•
Mixer motor is blocked.
•
Mixer motor driver on rack is board is faulty.
•
Check mixer motor belt and motor belt fixing.
•
Check if the pulleys on the motor and mixer rack are tightened properly and not slipping
E68 is the error code and the text in the status line Mixer motor time out. The mixer motor current exceeded
the maximum mixing current setting during mixing.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
165
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...