Work instruction Number
227
Page
1 of 1
Purpose:
Preparation for temporary storage and starting operation
after storage
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
002, june 2014
Preparing ESR Unit for temporary storage
1. Perform "End-of-day wash"
2. Remove all reagent containers and replace with containers with DI water
3. Start "Wash all pipettes"
4. Prime the diluter 5 times
5. Disconnect and remove reagent containers
6. Perform "Prime all" until all liquid is removed from all tubing
7. Remove tubing from reagent pumps
8. Release tubing from pinch valves (Diluter valve, Sample valve and Vacuum valve)
9. Check if the separator is clean
10. Clean outside of the sample needle
11. Clean outside of instrument. Wipe outer surface and stainless steel plate below the pipettes with disinfectant.
12. Disconnect waste connection
13. Disconnect power
14. Cover machine with dust protection
Before starting normal operation ESR unit
1.
In case of physical removal of instrument: check levelling
2. Check tubes and replace tubes if needed
3. Connect saline and rinse pump tubes
4. Re-install tubes in pinch valves
5. Replace fill nozzle o-ring
6. Check needle condition
7. Connect/check reagent containers, check expiry date of reagents
8. Connect power
9. Perform all prime steps separately and check liquid flow
10. Perform Fill and clean
11. Check if pipettes are leaking, replace check valves if leakage is persistent
12. Check sensors and vacuum pressure settings
13. Check instrument settings
14. Check communication interface settings
15. Perform aspiration with test blood before normal operation, check if all steps are performed without errors
Work instruction Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
221
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...