If there is no liquid flow and no reagent alarm:
1. One of the de-ionized water lines may be blocked or kinked.
2. Add one or drops of Starrsed Saline or Diluent to get some conductivity in the De-ionized water.
15.6.2 - Disinfectant
To disinfect the ESR analyser unit waste line, select from
Maintenance -> Prime / Clean
-> [Prime Disinfectant]
, the
vacuum pump starts, and sufficient liquid must be seen flowing through the thin tube connected to the side
of the rinse nozzle
After each wash cycle, approximately 0.5 ml of disinfectant will be flushed through the wash station.
If no disinfectant flows;
1. Unscrew the cap of the disinfectant bottle or cubitainer to check.
2. One of the disinfectant lines may be blocked or kinked.
15.7 - Needle system
As soon as the barcode is accepted or the ID number keyed in manually, the sample will be processed.
15.7.1 - Needle not in top position
E4
E4 is the error code and the text in the status line
Sample probe not in top position ! (home). Piercing
needle returns not at the home position after sampling a tube.
•
Check the home sensor.
•
Sample probe motor is faulty.
•
Sample probe motor driver on needle board is faulty.
•
Sample probe is blocked
•
Fatal error, call distributor.
E14
E14 is the error code and the text in the status line
Outer needle motor position error ! (home). Outer
needle did not reach the home sensor within a certain time limit.
•
Check home (top) sensor.
•
Outer needle (tube) motor is faulty.
•
Outer needle motor driver on needle board is faulty.
•
Motor is blocked.
•
Outer needle is blocked.
•
Fatal error, call distributor.
Troubleshooting
152
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...