Save settings
.
Perform the following steps to test the position motor:
1. Go to
IN position
: This is the position where the rack clamp is open. The rack can be inserted manually
in the clamp.
2. Go to
MIX position
: This is the rotation position where the rack clamp is engaged to secure the rack for
mixing the sample tubes.
3. Go to
(1)TUBE position
: This is the first hole position in the grabber cover plate at the needle position.
4. Go to
RELEASE position
: This is the position for releasing the rack. The grabber will be opened and the
rack is transported to the Rack out position.
Go to position
: Service purpose only (This is to move from the home position to any distance in millimeters
from the home position).
The following steps are needed for checking the barcode.
Take a sample tube with barcode and a rack.
1. Go to the
IN position
.
2. Go to
Mix position
.
3. Go to
Tube position number
1.
After pressing Read barcode
the barcode is shown above the button. If barcode label is not read correctly
see section Troubleshooting.
Switch
Check tube detection sensor
:
Use this switch to check the correct adjustment of the optical sensor that checks for the presence of sample
tubes in the rack. See “
Troubleshooting/ Tube sensor does not detect tubes
” for further instructions.
Switch
Always read barcode
:
Use this switch to read the barcode at every position in the rack. This tube is correctly logged in the rack,
even when the tube sensor is not detecting a tube. If the tube sensor is not detecting a tube and a barcode
could also not be read, the tube position is logged as empty.
See Service manual for further instructions
This section is used to make sure the needle sting position is centered in with the hole in the top bracket.
Perform the following steps for checking the needle sting location:
1. Go to
Tube position number
1.
2. Press outer needle
DOWN
.
3. Check the needle position.
4. Continue to press the needle
DOWN
and visually check the needle goes through the hole in the top of
the bracket.
5. Press outer needle
UP
.
If the needle is not passing the hole, see
.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
97
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...