1. Blocked filter. Replace blue disk filter.
2. Waste separator wrongly assembled or blocked.
▪
.
3. Main vacuum pump defective.
▪
Fatal error, only allowed
for a trained technician
.
4. Bad vacuum. Adjustment needed on the vacuum.
▪
Fatal error, only allowed
for a trained technician
.
5. Blockage in the 3 way vacuum manifold.
▪
Fatal error, only allowed
for a trained technician
.
15.6 - Flushing liquids
After each sample aspiration the entire system is washed automatically.
If there is no liquid flow:
•
Check that the peristaltic pumps are running. If the pump tubes are worn or leaking, replace the
tubes.
•
Check that the pump tubes are installed correctly.
•
Check the tubes between the containers and pumps/valves.
•
Unscrew the cap from the bottle or cubitainer. Check the pick-up tubes in the bottle or cubitainer.
•
Check the tubes for blockages or kinks.
15.6.1 - De-ionized water
Select from
Maintenance -> Prime / Clean
-> [Prime De-ionized water]
, the vacuum pump should operate and liquid
flows through the thin tube connected to the side of the fill nozzle cap.
After each sample aspiration, the fill nozzle is washed automatically with de-ionized water.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
151
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...