This screen shows all patient results that have been measured after the selected QC result and up to the
following QC result. The results are presented in the layout of the "
" screen. Depending
on the frequency of QC samples, related patient results may span over multiple days and are listed per date.
All general ESR data and errors of QC samples are shown here.
6.3.5.6 - Linked QC IDs
Use this screen to link the Starrsed Control sample ID with a Lab ID or to check which links are active.
1. "Starrsed Control sample ID": Enter the lot number or scan the barcode from the original Starrsed
Control tube label. If the original label is already covered by the Lab ID label, find the lot number and
barcode on the package insert.
2. "Lab ID": Enter the patient number or scan the barcode from the label that the lab is using to identify
the sample.
3. Click button "Link IDs" to add the linked ID's to the list. The "Date linked" will be added automatically.
4. Attach the Lab ID label on the Starrsed Control sample tube so that the original barcode is completely
covered to ensure that only the Lab ID barcode can be scanned by the Starrsed RS.
If the Starrsed Control sample ID is not correct or the expiry date is exceeded, a message will be shown and
the ID's are not added to the list.
To remove a link that will no longer be used, select the link in the table and click on "
Remove link
".
Depending on the optional setting "
Automatically remove linked QC ID after result
",
the
links can be removed automatically when a usable ESR result has been reported for this particular Lab ID.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
57
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...