15.7.2 - Sample probe fails to go down
Under normal circumstances, the sample probe goes down.
If sample probe fails to go down check the following:
•
Sample probe depth wrong. Check the correct sample depth:
Settings - General settings
-
Sample probe
depth
▪ If the sample probe has being set too deep, it will touch the bottom of the sample tube. The
sample probe then pushes the sample tube slightly downwards, and the aspiration cycle will be
aborted.
▪ A broken outer needle may cause a similar fault.
•
Check for mechanical obstructions.
•
Electronic failure.
▪ Fatal error, call distributor.
•
Mechanical failure.
▪ Fatal error, call distributor.
See the
for other sample probe errors.
15.8 - Air bubbles
After a normal aspiration, the Westergren pipette must be free of air bubbles. In the following examples
different patterns of air bubbles which can appear in the pipettes are shown. Air bubbles can affect the
sedimentation and are mostly reported as errors and no ESR result is reported.
Usually bubbles are caused by a leakage at the bottom of the pipette. If air bubbles are visible in the pipette,
check the following :
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
153
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...