E94: Position motor error! Not on
position! (MIX)
Gripper did not reached the MIX
position after reading the barcode.
Starting point: FILL position.
See E61
E95: Position motor error! Not on
position! (FILL)
Gripper did not reached the FILL
position after mixing the rack.
Starting point: MIX position.
See E61
E96: Position motor error! Not on
position! (FILL)
Gripper did not reached the FILL
position. Starting point: last
sampled tube.
See E61
E97: Position motor error! Not on
position! (IN)
Gripper did not reached the IN
(catch) position. Starting point:
MIX position.
See E61
E98: Position motor error! Not on
position! (IN)
Gripper did not reached the IN
(catch) position after gripper error.
Starting point: MIX position.
See E61
E99: Position motor error! Not on
position! (MIX)
Gripper did not reached the MIX
position. Starting point: FILL
position.
See E61
E100: Position motor error! Not on
position! (MIX)
Gripper did not reached the MIX
position. Starting point: IN
position.
See E61
E101: Position motor error! Not on
position! (FILL)
Gripper did not reached the FILL
position. Starting point: last
sampled tube.
See E61
E102: Position motor error! Not on
position! (MIX)
Gripper did not reached the MIX
position. Starting point: FILL
position.
See E61
E104: Needle unit not in up
position!
Could not start the position motor,
because the outer needle or
sample probe is not its home
position (top).
•
Check outer needle home
sensor.
•
Check sample probe home
sensor.
•
Faulty Outer needle motor.
•
Faulty Sample probe motor.
•
Check if needles are
blocked.
•
Faulty motor drivers on the
needle board.
E116-118
Quality Control Errors
Appendix for Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
195
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...