We recommend that this procedure is carried out by qualified service engineers.
Detailed instructions can be found in:
13.6 - Level 1 maintenance
Annual maintenance (Level 1 maintenance) is Level 2 Maintenance and the following extras;
We recommend that this procedure is carried out by qualified service engineers.
The following items need to be replaced annually:
1. All tubing
ESRI079200 and additional tube set
.
2. Waste pump motor
ESRI090920.
3. Waste pump cassette
ESRI090921
.
4. Blue Vacuum filter disc. Part no
QWLV040003
.
5. Fill block washer. Part no
ESRI030906.
6. Waste container filter disc
QWLV040001.
(only applicable if internal waste container is used)
7. Teflon tip of syringe on the Diluter assembly (*Only applicable for Diluter syringe QWLV030009, Diluter
syringe QWLV030021 has to be replaced after 2 years).
The following items need to be replaced annually:
1. Outer needle and sample probe (if needed)
2. Pipette valve bodies (84 pieces)
QTST040001
.
Check:
1. Adjustment of fill nozzle and rinse nozzle.
2. Check measurement unit with Control pipettes.
All these parts (except for the Control pipettes) are included in the PM-kit (Periodical Maintenance Kit).
See Appendix - Maintenance schedule
13.7 - Two years maintenance
Since 2020 a 2-years maintenance kit is available: order numer SESRI120906.
This kit contains two parts of tubing of the vacuum regulator and a diluter syringe and applicable O-ring.
13.8 - Reagents replacement
The Starrsed RS can be used with the small onboard reagent containers or with the Mechatronics bulk
reagent containers if long level sensors are available.
General maintenance
MRN-071-2-EN version 7.02 Starrsed RS
137
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...