12
ESR analyser unit system messages
The ESR analyser unit generates four main types of error messages;
•
System messages.
•
Test messages.
•
System time-out messages.
•
Error messages.
12.1 - System messages
During normal operation the following "
System messages
" may occur:
1.
Waiting tube
▪ If a filled pipette is at the measuring position before the elapsed time has finished and the operator
is ready to fill the next pipette, the
Waiting tube message will be displayed.
▪ To continue the sample loading sequence the operator must wait until the pipette at the measuring
position has been measured.
2.
Reagents level empty message
▪ All reagent bottles or cubitainers have level detectors; the display shows an error that indicates
which reagent bottle(s) or cubitainer(s) is (are) empty.
▪ The expiry date of the reagent is exceeded or the packaging is opened longer than three months.
▪ Prepare new reagent as described in section
3.
"Waste bottle full" message or "No waste bottle" message
▪ The waste container also has a level detector. If a waste error is shown on the display, the Starrsed
RS will stop filling and cleaning pipettes until a new or empty container has been installed.
▪ Empty and/or replace waste container and press Clear error.
4.
Fatal separator error
▪ The Separator container has a level detector. If the "
" message is indicated on the display, the
ESR analyser unit will stop the "rinse" cycle until the separator is empty.
▪ The cause of this problem can be foam or the waste pump is not working. The ESR analyser unit
will continue with the measurement and sends the ESR results on time, but the rinse and fill
sequences are stopped until the error is solved.
12.2 - Test messages
During the start-up sequence all the positioning sensors are tested, if incorrect the instrument will generate
a
Test message
,
ESR analyser unit system messages
128
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...