▪ Select Set rinse position.
▪ Change Rinse position.
▪ Enter the correct value of the pipette number at the rinse station.
▪ Close pop-up menu.
2. Select Set rinse position
▪ Change Rinse position
▪ Press Learn carousel position
▪ When all positions are learned, close pop-up menu.
Check if the position error is solved: Select
Go to rinse position
, enter a higher value than the actual rinse
position and the carousel should move without position error.
If position error is solved:
1. Leave service mode.
2. If samples or liquid is present in any pipette, go to tab
Maintenance - Prime/clean - Wash all pipettes
which
removes the liquid from the pipettes.
All sample information will be lost.
If there is still a position error:
If the position is still not correct or if position errors occur frequently and the solutions above do not solve
the problem, the following procedure must be carried out:
1. Leave Service mode
2. a) If liquid or samples are present in isolated pipettes, use a combination of
Go to rinse position
at the
service tab and
Maintenance -Prime rinse solution
to remove the blood samples. All sample information will
be lost.
or
b) Go to tab
Settings
and select
Delete pipette data
. The carousel stops moving. Use the sequence
mentioned above to remove the blood samples.
3. Check mechanical connection potentiometers.
If the position error is not solved after all the mentioned actions, the position device needs replacement.
This has to be done by a trained technician, call distributor.
15.15 - Separator error
If it takes too long for the waste pump to empty the waste separator, the system generates a separator
error.
Separator error may be caused by:
Extensive foam build-up in the waste
separator.
Check the separator assembly and connections for possible
air leaks.
Waste-tube between waste separator and
waste pump is blocked.
Replace the tube.
Troubleshooting
MRN-071-2-EN version 7.02 Starrsed RS
161
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...