Work instruction Number 187
Page
1 of 1
Purpose:
Daily maintenance
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
002, March 2019
Prepare disinfectant:
(if not already prepared).
This disinfectant is for cleaning of all external parts that are exposed to blood.
1. Go to tab
[Maintenance]
and perform the
[End-of-day wash]
procedure.
2. Check system for leakage.
▪ Inspect the peristaltic pump tubes and connections for leaks.
▪ Check that liquid does not run back after the pumps have stopped.
3. Clean the outer needle with disinfectant
4. Check tubing from the syringe for trapped air bubbles.
5. Check Diluent syringe for trapped air bubbles.
6. If trapped air bubbles are found, go to tab
[Maintenance],
click button
[Prime / Clean]
and perform the
[Prime
Diluent/Diluter]
function.
7. Wipe outer surface and stainless steel plate below the pipettes with disinfectant.
Work instruction Starrsed RS
224
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...