15.8.1 - Foam in column
A layer of air bubbles that is concentrated on top of the blood column does not
affect the sedimentation process itself. The sedimentation develops normally
below the bubbles. However, too many bubbles bring about a shortening of the
effective blood column, which is a deviation from the Westergren method.
A layer of bubbles up to 5 mm: No message. Normal ESR result is reported.
A layer of bubbles from 5 to 25 mm: ESR warning 6: "Bubbles on top". Results
should be reviewed before release.
A layer of bubbles larger than 25 mm: ESR Error 3: "Too many borders found".
No ESR result is given.
1. Check that tube connections are not leaking.
2. Check the fill nozzle condition:
▪ Inspect for any cracks or deep scratches in the base that holds the fill
nozzle washer or O-ring.
3. Check for air in diluter system.
4. Check that the sample probe O-ring is not leaking.
5. Check transparent mixing- piece block for cracks.
Troubleshooting
154
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...