1. Blood clots or rubber debris from the tube cap in the sample.
▪ Check the condition of the outer needle.
▪ Check the mixing time of the tube
Settings
-
General settings
-
Mixing interval
<default 90 sec>
2. Filling procedure stopped by operator.
3. Insufficient sample volume.
▪ Should be at least 1.4 ml.
4. Faulty fill nozzle or fill nozzle washer/O-ring.
▪ Check fill nozzle and washer/O-ring.
5. Incorrectly adjusted sample probe depth.
▪ Check needle depth
Settings
-
General settings
-
Sample probe depth
<default 5 mm>
6. No or poor vacuum.
▪ Check vacuum
Maintenance
-
Check sensors
-
Check flow sensor
15.14 - Carousel position error
E18 is the error code for Carousel position error. Starrsed RS was not able to position the carousel. There
was a difference found in the pipette memory position table and the actual measured position of the position
sensor.
A position error can occur if the Starrsed RS is switched OFF while the carousel is moving to the next
position or when the Rinse nozzle sticks up.
Warning
: In case of the uncertainty about the position and linked patient data, is it highly
recommended to wash all samples after a position error.
Position error solutions
1. Check in tab
Settings
(password XXXXXX)- Carousel control the Rinse position.
Troubleshooting
160
MRN-071-2-EN version 7.02 Starrsed RS
Содержание STARRSED RS
Страница 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Страница 2: ......
Страница 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 35: ...5 3 E C Declaration Starrsed RS Standard Operating Procedures S O P MRN 071 2 EN version 7 02 Starrsed RS 35 ...
Страница 84: ...Flow Absolute flow test Leakage test Starrsed program 84 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 85: ...Fillsensor air flow test Fillsensor glass flow test Starrsed program MRN 071 2 EN version 7 02 Starrsed RS 85 ...
Страница 172: ...16 Appendix for Starrsed RS Appendix section Appendix for Starrsed RS 172 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 175: ...16 2 Cable connection diagram Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 175 ...
Страница 176: ...16 3 Tube connection Starrsed RS Appendix for Starrsed RS 176 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Страница 200: ...For manual patient id handling Beckman Coulter Appendix for Starrsed RS 200 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Страница 202: ...Bayer Horiba Yumizen H Horiba ABX Appendix for Starrsed RS 202 MRN 071 2 EN version 7 02 Starrsed RS ...
Страница 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Страница 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...